This Fluoride calibration standard solution is designed to be used with Horiba LAQUAtwin F-11 ion fluoride meter.

Fluoride meter needs to be calibrated daily, or every 10 measurements, depending on your frequency of testing.

This set contains 6 x 14 ml bottles and will last for approximately 400 calibrations.

 

How to calibrate LAQUAtwin F-11 fluoride meter (manufacturer recommendations):

 

1) Calibrate with 1:1 mixture of TISAB solution and standard solution

2) Add 1ml TISAB solution to the beaker

3) Add 1 ml of 10 pm standard solution to the beaker

4) Add this mixture to the sensor area, press 'CAL' 2 times and wait for the smiley face to appear.

 

Calibration is now completed with the first calibration point. 

Repeat steps 1-3 with mixing TISAB solution and 1 ppm standard solution for the second calibration point.

Fluoride Calibration Solution 10 ppm 514-F-10